ISSN 2063-5346
For urgent queries please contact : +918130348310

Bioanalytical reverse phase high performance liquid chromatography method development and validation for the estimation of Tetrabenazine in rat plasma

Main Article Content

Shashi1, Yadav Sarvi Rajesh1, Bala Vikash R1, Kusman Sekhar Behra1, Narendra Kumar Pandey1, Bimlesh Kumar1, Sachin Kumar Singh1, Dileep Singh Baghel1, Ritu Gupta2, Saurabh Singh1, Kalvatala Sudhakar1
ยป doi: 10.48047/ecb/2023.12.5.153

Abstract

Tetrabenazine (TBZ) in rat plasma can be estimated using a reverse-phase high-performance liquid chromatography (RP-HPLC) approach employing benzoquinolizine as an internal reference. ICH M10 criteria were used to validate the created approach. Acetonitrile (ACN) and 0.1% formic acid were used in a ratio of 90:10 v/v for the separation, with a flow rate of 1 mL/min and a detection wavelength of 283 nm. Benzoquinolizine and TBZ had retention times (RT) of 3.665 and 5.064 minutes, respectively. The developed method's regression coefficient (r2) of 0.996 revealed that it is linear in the 50โ€“250 ng/mL range. Indicating their accuracy and precision,

Article Details