ISSN 2063-5346
For urgent queries please contact : +918130348310

DEVELOPMENT AND VALIDATION OF SIMPLE, SELECTIVE AND SENSITIVE LIQUID CHROMATOGRAPHY TANDEM MASS SPECTROSCOPY METHOD FOR QUANTIFICATION OF DOLUTEGRAVIR IN HUMAN PLASMA

Main Article Content

Nameera Jabeen, Dr. M. Akiful Haque,
» doi: 10.31838/ecb/2023.12. 4.124

Abstract

A simple, rapid, sensitive and novel technique for quantifying of Dolutegravir in human plasma was developed and validated. A protein precipitation technique was used as sample preparation method and Tenofovir is employed as Internal standard (IS). Chromatographic condition was conducted using a Agilent, Zorbax, XDB C18 2.1 × 50 mm, 5 µm., 5 Micro meter, and a Mobile Phase (70: 30) of proportion Acetonitrile;1% Formic acid delivered at 1.20 ml/min and injection volume of 10 µL with run time of 3 minutes. ESI-MS with positive ion (PI) mode at atmospheric pressure was employed to carry out detection. The validated Linearity range is 2ng to 1000ng/ml with correlation value of 0.998 , excellent accuracy and precision was obtained. The retention time of Dolutegravir and Tenofovir 1.37 and 1.20 min.

Article Details