DEVELOPMENT, VALIDATION AND FORCED DEGRADATION FOR QUANTIFICATION OF TECOVIRIMAT IN BULK AND DOSAGE FORMS BY UPLC-MS/MS

Main Article Content

Anusha Kota1, Dr. M. V. Kumudhavalli*2, Dr. B. S. Venkataswarlu3, Dr. Margret Chandria4, Dr. J. S. K. Nagarajan5, Ramu Samineni6
» doi: 10.48047/ecb/2023.12.5.021

Abstract

UPLC has been used to find a selective and new way to test samples of Tecovirimat in bulk and formulations.The analytical method used phosphate buffer (pH: 7.0): acetonitrile, (75:25 %v/v), the cortecs C18 column and the analysis took five minutes.The flow of mobile phase through the column was 1 ml/min. The size of the sample was 10 μL. The detection was done with a photo diode array detector at a wavelength of 220 nm. At a retention time of 6.22min, Tecovirimat performed well. With a linear range of 1-1000ng/ml, the curve's r2 value of 0.9997 shows a very high degree of correlation.