Volume - 13 | Issue-1
Volume - 13 | Issue-1
Volume - 13 | Issue-1
Volume - 13 | Issue-1
Volume - 13 | Issue-1
Objective: To develop and validate economic, user friendly, precise, specific and accurate RP-HPLC method for Acamprosate calcium. Methods: The RP-HPLC method uses Reverse phase C18 column (Comosil) (4.6 X 250mm, 5μm), a mobile phase of acetonitrile, (0.5%) at flow rate 0.7 ml/min, at 25oC ± 2oC and detection wavelength was set at 210 nm. Results: The developed method of RP-HPLC displayed linearity in the range of 2.9 to 22.6 part per million. The specificity obtained indicated non-interference from other impurities. The LOQ for acamprosate calcium obtained to be 39:1 and LOD for acamprosate calcium obtained to be 13:1.